Calibration register
The ledger that lists every measuring instrument on site with its control number, range, calibration interval and the date it was last calibrated. The next due date is calculated from the interval, so a glance down the column shows what is coming up. It supports the traceability requirement of ISO 9001 and the thermometers and scales behind your HACCP and inspection records.
Fill it in right here
Click any blank cell and type. Totals and amounts calculate as you go, and you can print it, save it as PDF or download the filled-in Excel file.
My documents
Saved in this browser only. Save versions to keep separate customers or dates. Clearing site data removes them; download important files.
| Control no. | Instrument | Model / maker | Serial no. | Measuring range | Interval (months) | Last calibrated | Next due | Calibrated by | Result | Location | Certificate no. / notes |
|---|---|---|---|---|---|---|---|---|---|---|---|
Next due = last calibration date + interval in months (Excel EDATE). Intervals come from your quality manual, the manufacturer or the calibration lab and are reviewed against drift history. A conditional format on Next due (earlier than today) flags overdue instruments.
An instrument found out of tolerance is taken out of use and tagged, and the validity of earlier measurements made with it is assessed (ISO 9001:2015, clause 7.1.5.2); record the outcome in a nonconformance report.
How to use this tool
- Enter the company, department, last update date and the person responsible in the header; they are written into the file.
- Open the Excel file and add one row per instrument: control number, name, model, serial number, range, interval in months and the last calibration date. The next due date is calculated.
- After each calibration record the lab or in-house standard, the result and the certificate number in the notes, update the date and file the certificate with the register.
Good to know
- Intervals are not built into the file: use the values from your quality manual, the manufacturer or the calibration lab. ISO 9001:2015 clause 7.1.5.2 (measurement traceability) is the reference behind this register.
Frequently asked questions
What should a calibration register contain?
At least a unique identification (the control number), the instrument name and model, serial number, measuring range, calibration interval, last and next due dates, who calibrated it, the result and where the instrument is used. ISO 9001:2015 clause 7.1.5.2 asks that measuring equipment is identified so that its calibration status can be determined; this register and the label on the instrument do that job.
How is the next calibration date calculated?
The next due date is the last calibration date plus the interval in months, using the EDATE function: 12 months after 15 March is 15 March of the following year, and a date at the end of a long month moves to the last day of a shorter one. If the interval or the last date is empty the cell stays blank. To flag overdue instruments, add a conditional format that highlights dates earlier than today.
How often must instruments be calibrated?
There is no single answer. The interval depends on the recommendation of the manufacturer, how often and how harshly the instrument is used, the accuracy needed and its drift history, so check your quality manual or the customer requirements. Many sites start with 12 months, shorten the interval for instruments found out of tolerance and extend it when results stay stable.
What do I do when an instrument is found out of tolerance?
Take it out of use and tag it, then assess whether measurements made since its last good calibration are still valid; ISO 9001 requires that assessment, and product released on those readings may need to be re-checked. Record the result as out of tolerance, adjust or repair and recalibrate, and document the outcome in a nonconformance report. The adjusted option marks instruments corrected during calibration.
How do I record internal and external calibration?
Write the name of the calibration laboratory in the calibrated-by column for external calibration, and the master standard used for internal calibration. Put the certificate number in the notes column and file the certificate with the register. For traceability the standard or lab should itself be calibrated against national or international standards, ideally by a laboratory accredited to ISO/IEC 17025.
Templates are starting points. Adapt the fields and limits to your own procedures and the rules that apply to you.